Key Features

  • 5 in 1 composite analyzer scanner (AC / DC/ IR / IWT / DCR)
  • 5kV AC/6kV DC Hi-pot test
  • 5kV Insulation Resistance test
  • Impulse Winding Tester (IWT)
  • IWT high sampling rate(200MHz)
  • 10 channels 4-wire DCR test
  • Δ /Y motor DCR calculation
  • HSCC (High Speed Contact Check)
  • Support max. 40 channels scanning test
  • English, Traditional Chinese and Simplified Chinese User Interface
  • USB waveform storage& Hand copy function
  • Graphic color display
  • Standard LAN,USB,RS232 interface
  • GFI (Ground Fault Interrupter) for bod protection

Chroma 19036 is the industry’s first test device that combines the functions of impulse tester and hipot analyzer for testing the impulse of wound components. The tester has 5kVac/6kVdc high voltage output and 6kV impulse voltage that can comply with the wound components test demands by providing maximum 10 channels output for multichannel scanning tests to save time and labor costs.

The quality verifications of wound components include AC/DC hipot test, IR test and impulse winding test. Chroma integrates the above tests into 19036 Wound Component EST Analyzer that can perform safety tests on wound products like motors, transformers and heaters to verify their quality with efficiency.

Since the poor insulation of coil often causes layer short, cross-line short and pin short, layer short circuit test is required for coils as the reason could be initial design error, poor fabrication process or bad insulation material. Moreover, the wound components for safety tester need to be tested with Impulse Winding Tester (IWT) to check the insulation ability of windings. It can measure multiple test points in one test instead of switching test points manually.

Combining with impulse winding test function the 19036 has 6kV impulse voltage, AREA SIZE COMPARISON, DIFFERENTIAL AREA COMPARISON, FLUTTER DETECTION and LAPLACIAN DETECTION judgment that are effective methods for detecting poor coil insulation.

19036 is equipped with a patented 4-wire test port that has both Drive and Sense in compliance with hipot specification to provide 10 channels of 4-wire test functions. Up to 40ch of scanning test can be conducted when 19036 is configured with 16ch scan box.

19036 also has HSCC functions to check for any bad contact prior test. It can solve the test fail problems caused by motor or transformer bad contact and improve the test quality as well as prolong the test equipment life.

The motor test standard such as UL 1004-1 requires high power safety tester. For the user that needs to test large leakage current or perform large equipment electrical safety tests, Chroma 19036 that has the capability of outputting and measuring AC 100mA/ DC 20mA with high power hipot tests and other safety tests integrated into one is the most suitable device to bring the maximum benefit to production line and quality assurance. The 500VA design is also compliant with IEC/UL for output power requirements.
 

Measurement Technology

HI-POT Test Flashover Detectionarc
Fast transient on detected leakage current is defined as flashover in hi-pot test. The 19036 has Flashover Detection as other Chroma safety test series. Flashover is nonsequential discharge generated by the inner or surface of insulation component which cause DUTs to lose original insulation feature, generated carbonation electrically conductive path and damages the product. As flashover cannot be detected by leakage current rms value, Flashover Detection is one of the inevitable inspection items in HV test.

Leakage Current

Impulse Testing Overview
The impulse winding test is to impose a non-destructive, high speed and low energy voltage impulse on the DUT (Device Under Test) to analyze/compare the equivalent waveform of yield and defect products for good and no good judgement. The main function of impulse winding test is to discover the potential defects such as layer short, corona or partial discharge that is difficult to find in wound components in the early phase.
Four Kinds of Waveform Judgement for Testing

  • Area size
  • Differential area
  • Flutter value
  • Laplacian value

Use Laplancian for calculation can detect the waveform discontinuity caused by electrical discharge effectively.

Impulse Testing Overview

 

DCR Measurement

  • DCR Measurement (two-wire/four-wire)
    Chroma 19036 can choose to use 2-wire or 4-wire DCR measurements. The new patented port design provides maximum 10 channels 4-wire DCR measurements for high accuracy tests on multiple wound samples like motors and transformers. The measurement range is from 0.1mΩ to 500KΩ.
  • DCR Balance
    When the DCR of the motor with 3 windings is unbalanced, the rotation will not be balanced as well and it could cause bad quality for long term use. The judgement of DCR balance is to subtract the maximum and the minimum value of winding and the product is no good if the value exceeds the set range. It is a reliable assistant tool for testing motor products.
  • Temp Compensation
    When measuring a smaller DCR, it often encounters the problem of temperature difference. The measured resistance is varied with the temperature. Thus, the temp compensation function is included in the 19036. Through the conversion of temperature coefficient, it changes the DCR to the measured value under standard temperature to reduce the influence caused by temperature difference.

High Speed Contact Check HSCC
If an open occurs to electrical safety testing circuit it may cause a defected product to be judged as a good one. The contact check can detect and screen the products when short circuit occurs to reduce the damage of fixture device and save the test cost. The HSCC function can scan rapidly if the DUT circuit contact is correct . This new technology allows the contact check before hipot test to be done faster than ever. The 19036 also has High Frequency Contact Check (HFCC) and Open Short Check (OSC, Patent no. 254135) functions that can detect if there is Open (bad contact) or Short (DUT short circuit) between the winding and the iron core.

Substep for Testing Multiple Wound Components 
Parallel hipot test is often used to improve the manufacturing inspection speed; however, it is unable to set the current high/low limit correctly when doing a parallel test and it causes the leakage of defected products or misjudging the good to no good. It also adds workstations for testing the defected products and increases the cost. The 19036 offers the sub-step function to solve the trouble caused by the parallel test. When the parallel test is required for production, it programs fail as the sub-step launch condition which means only when the main test item (parallel) fails the sub-step test item (single) will be performed to judge the defect DUT. The implementation of this function can optimize the productivity and quality inspection.

GFI Human Protection 
The ultimate purpose of safety testing is to protect users from electrical hazard.The19036 also has GFI function to protect the operator. The GFI function can cut off the power output immediately while the human body is suffering electric shock. The GFI function detects the current from earth GND (Ioperator) and LOW terminal (Idevice). The voltage output will be cut off if the current is over 0.5mA.

GFI HUMAN PROTECTION

Product Applications
From EV motors, servomotors, ascension motors to fans all rotary motor products need to perform impulse test, hipot test and DCR measurements to ensure the product quality also refer to JB/T 7080 GB Machinery Industry Standard for testing. The DCR measurement of the 19036 is in 4-wire type and each terminal has Drive and Sense total 10 independent channels for scanning test 3 DUTs at once to increase productivity. Each channel can be set to Hi/Lo/Open respectively.

Y-type motor for example, the test itemsY-type motor for example, the test items are:

  • HSCC / OSC
  • DCR Test
  • Impulse Test
  • Hi-pot –Sub step test

 

△-type and Y-type motor△-type and Y-type motor
To solve the problem of unable to do DCR measurement on the △-type and Y-type motor winding (no center-tapped) directly, Chroma 19036 has added △-type and Y-type motor winding DCR calculation functions to obtain the value of R1,R2 and R3.

 

40 Channels scanning test40 Channels scanning test
A190359/A190362(4-wi r e) s c anne r ha s 16 tes t channel s and each of them can be set to H (High voltage, Hi), L (High voltage, Lo) or X (Open). The combination of 19036 and A190359/A190362(4-wire) can apply to multiple PINs or small amount but diversified DUTs as well as cell type production line to complete all tests within one station.


 



Model
Description
19036 Wound Component EST Analyzer
A190359 16ch HV External Scanning BOX